Power Monitoring & DAQ

Yokogawa GS820 Multi Channel Source Measure Unit
 
 
Product Overview
Application Examples
  • DC voltage/current reference, pulse voltage/current source
  • V-I curve trace, semiconductor parametric test
  • Programmable arbitrary waveform generation
  • Electric load, resistance measurement
  • Production tes

Description

The GS820 is equipped with two analog channels with each channel consisting of a constant voltage source VS, a constant current source IS, a voltmeter VM, and an ammeter IM. The two source measure channels are isolated.

Four-quadrant operation consisting of source operation (current source) and sink operation (current sink) is available with ranges up to 7 V and 3.2 A or 18 V and 1.2 A. The output and measurement resolutions are 5.5 digits. Other features include:

  • Isolated 2-channel source and measurement function
  • Source and measurement ranges:

18V model: 7 V and 3.2 A or 18 V and 1.2 A
50V model: 20 V and 1.2 A or 50 V and 0.6 A

  • 1 pA resolution at extremely small current range 200 nA
  • Generate arbitrary waveforms consisting of up to 100,000 points at 100 μs intervals
  • Channel expansion through master-slave synchronization link
  • Fast test speeds
  • 16-bit digital I/O (model 765602/12)
Tm Gs820 02
2-channel display example
(256 x 64 dot matrix display)

Source and Measurement Range

Four-quadrant operation consisting of source operation (current source) and sink operation (current sink) is available. The output and measurement resolutions are 5.5 digits. Two models are available for your application.

18 V model (765601/02)
  • Voltage ranges:
    • 200 mV, 2 V, 7 V, and 18 V
  • Maximum output current:
  • ±3.2 A (at an output voltage of ±7 V or less) ±1.2 A (at an output voltage of ±18 V or less)
  • 200 nA, 2 µA, 20 µA, 200 µA, 2 mA, 20 mA, 200 mA, 1 A, and 3 A
  • ±18 V (at an output current of ±1.2 A or less) ±7 V (at an output current of ±3.2 A or less)
  • Current ranges:
  • Maximum output voltage:
Screen Shot 2015 01 29 At 3.24.38 PM 1

50 V model (765611/12)
  • Voltage ranges:
    • 200 mV, 2 V, 20 V, and 50 V
  • Maximum output current:
  • ±1.2 A (at an output voltage of ±20 V or less) ±0.6 A (at an output voltage of ±50 V or less)
  • 200 nA, 2 µA, 20 µA, 200 µA, 2 mA, 20 mA, 200 mA, 0.5 A, and 1 A
  • ±50 V (at an output current of ±0.6 A or less) ±20 V (at an output current of ±1.2 A or less)
  • Current ranges:
  • Maximum output voltage:
50 V Model

Specifications & Detail

GS820 Construction and Functions

The GS820 is equipped with two analog channels with each channel consisting of a constant voltage source VS, a constant current source IS, a voltmeter VM, and an ammeter IM.

The two source measure channels are isolated.


Source and Measurement Functions
  • Voltage source and current measurement (VS & IM)
  • Current source and voltage measurement (IS & VM)
  • Voltage source (VS)
  • Current source (IS)
  • Voltmeter (VM)
  • Ammeter (IM)
  • Resistance meter (IS&VM)

These functions can be selected for each channel to form an arbitrary combination of functions.

Allows voltage sensing of a two-wire system or four-wire system by switching between local sense and remote sense.

Construction & Functions

Combination of Source and Measurement Functions

The combination of the source and measurement functions of two channels allows the testing of various DUTs.

Yokogawa


Source and Measurement Timing

Basic Source Measurement Timing

The GS820 performs generation and measurement using its internal timer or a trigger input such as an external input signal. When a trigger signal is received, the GS820 starts generating a signal after the source delay time elapses and carries out a measurement after the measure delay time elapses over a given integration time. The measurement integration time can be set in the range of 0.001 PLC to 25 PLC.*² Additionally, the GS820 provides an auto zero measurement function, which measures the internal zero reference after the measurement and performs offset correction in real-time. The integration time of the auto zero measurement is equal to the measurement integration time setting.



Yokogawa

Timing Settings Using Various Trigger Sources

The GS820 allows the generation trigger source and measurement trigger source to be set separately. There are two types of constant period timers and an external signal input that can be used for the generation trigger source. In addition to these sources, source change point and sweep end point can be used for the measurement trigger source. Because the source trigger and measurement trigger can be set separately and also separately for each channel, source and measurement under various connection conditions and timing combinations can be accommodated. There is also an auxiliary trigger that can be activated using an external signal or a program event. The source delay, measure delay, and integration time can be set separately for each channel.



Yokogawa

Asynchronous Operation of Source and Measure

The various trigger sources available on the GS820 allow the source and measurement to be executed asynchronously. The figure below shows an example in which separate timers are used for the source and measurement to achieve multiple measurements in a source cycle.



Yokogawa




Sweep Function 1: Preset Sweep

The voltage/currentgeneration block of the GS820 operates in DC generation mode or pulse generation mode. Each generation mode has preset operation modes such as continuous output, linear sweep, and log sweep that allow the user to perform sweep operations by setting simple parameters. The output level can be changed at a minimum of 100-µs intervals*¹ in each sweep mode.



*1: Minimum program cycle

100 µs: If the measurement function is OFF

500 µs: If the measurement function is ON, two channels are used, and auto zero is OFF

Yokogawa

Sweep Function 2: Arbitrary Waveform Generation of Up to 100,000 Points and Simultaneous Sweeping of Control Parameters

In addition to the preset sweep functions described above, the GS820 is equipped with a programmable sweep function that allows the user to define the sweep pattern. A user can create or edit arbitrary waveform data (CSV format) of up to 100,000 points using a spreadsheet or text editor. The GS820 is also of capable of sweeping the timing and control parameters in addition to the source level. This allows a control sequence that is synchronized to the waveform generation timing. The sweep program can be changed at a minimum of 100-µs intervals*¹ in programmable sweep mode.

*1: Minimum program cycle

100 µs: If the measurement function is OFF

500 µs: If the measurement function is ON, two channels are used, and auto zero is OFF

The values do not include the hardware operation time corresponding to the control parameters.

Yokogawa



Yokogawa


Test Sequence Editing

(Application to Auto Testing Equipment)



The GS820 allows the editing of test sequences suitable for auto testing on the production lines. A user can write program fi le parameters that are vital to auto testing such as the source value, measured value, high limit for comparison, low limit for comparison, comparison result, control bit output, etc. Because the program fi le is in CSV format, a popular spreadsheet application can be used to edit and view the program.



Yokogawa



Yokogawa


Test Speed (Improvement in the Takt Time in the Production Line Test)

The GS820 provides fast operation for production line tests. The measured results of test speeds (reference data) are indicated below.

Yokogawa

Channel Expansion

Expansion up to 10 Channels Using the Master-Slave Operation



Multiple GS820s can be connected as shown below and used as a multi-channel source measure unit. The master-slave feature allows the program data of all connected channels to be set and collected by simply accessing the master unit. The master unit (1) distributes the source data to the slave units or (2) collects and merges the measured data of all slave units. Complete synchronization of all channels can be achieved by connecting the exclusive trigger signal line.



Yokogawa


Zero Generation Function of Voltage and Current

Fast Load Disconnection without Chattering



The zero generation function of the GS820 generates zero voltage or current as well as controls the current/voltage limiter to limit the load current. The GS820 stops applying the voltage or supplying the current to the load in the zero generation state allowing the DUT to be disconnected with the output relay turned ON. This function avoids the problems of chattering and contact life of the output relay and reduces the time for turning ON/OFF the output.


Yokogawa

Data Sheet & Downloads